Analysis of Catecholamines by Electrochemical Detector (ECD)

November 9, 2021

Introduction

 

Catecholamine is the generic term for the chemical compounds with catechol and amine groups and functions mainly as a neurotransmitter in vivo. Epinephrine (also called adrenaline), norepinephrine (also called noradrenaline), dopamine and their metabolites such as metanephrine and normetanephrine are known as catecholamines. An analytical measurement of catecholamines in the urine or blood is employed to examine and diagnose a variety of clinical conditions like high blood pressure, phaeochromocytoma and neuroblastoma.

The article shows the separation and subsequent detection of five types of catecholamines by high performance liquid chromatography (HPLC) connected to the GL Sciences’s ECD, ED743.

Experimental

[Conditions]
Column: Unifinepak C18 (4.6 mm I.D. x 150 mmL, 5 μm)
Eluent: 10 mM hexanesulfonic acid sodium and 20 mM citrate acid in water/acetonitrile (90/10)
Flow rate: 1.0 mL/min
Column temp.: 35 ºC
ECD condition: 1200 mV vs. Ag/AgCl (Diamond Electrode)
Injection volume: 20 μL
Standard: Epinephrine, Norepinephrine, Dopamine, Metanephrine and Normetanephrine (Refer to the each data for concentrations.)

[Structures]

Keywords

Catecholamine, Unifinepak C18, HPLC, Electrochemical Detector

Results

Figure 1 shows the chromatograms of catecholamine standard samples. Five types of catecholamines were very well separated and detected.

Figure 1 Chromatograms of Catecholamine Standard Samples (Each, 5-100 ng/mL)
1: Norepinephrine, 2: Epinephrine, 3: Normetanephrine, 4: Dopamine, 5: Metanephrine

Table 1 shows the calculation results of linearity, reproducibility, detection limits and limit of quantitation of catecholamine standard samples, while Figure 2 shows the calibration curves of these samples. In every type of catecholamines linearity, reproducibility and detection sensitivity were confirmed.

Table 1 Linearity, Reproducibility, Detection Limits and Limit of Quantitation

Types of Catecholamines Linearity (r)*1 Relative Standard Deviation [%]
(n = 6)*2
Detection Limits*3 [pg]
(S/N = 3)
Limit of Quantitation*3 [pg]
(S/N = 10)
Retention Time Peak Area
Norepinephrine 0.9999 0.08 1.59 31.1 103.7
Epinephrine 0.9999 0.07 1.55 61.7 205.7
Normetanephrine 1.0000 0.06 1.11 42.6 142.0
Metanephrine 0.9999 0.11 1.14 44.8 149.5
Glucose 0.9999 0.08 0.70 70.4 234.7

*1 Range of calibration curve: 5-100 ng/mL each
*2 Calculated from the measurement result of each 25 ng/mL standard solution
*3 Calculated from the measurement result of each 5 ng/mL standard solution

Figure 2 Calibration Curves of 5 Types of Catecholamines (each 5-100 ng/mL)

Figure 3 illustrates as reference data the chromatogram of a standard sample of catecholamine detected by a UV detector (UV-4075) under the same analytical condition. Table 2 shows the sensitivity comparison of an ECD and an UV detector. The ECD has proved to have 100 times higher sensitivity than the UV detector for all of 5 types of catecholamines.

Figure 3 UV Chromatography of Catecholamine Standard Samples (each 1000 ng/mL, 280 nm)
1: Norepinephrine, 2: Epinephrine, 3: Normetanephrine, 4: Dopamine, 5: Metanephrine

Table 2 Sensitivity Comparison of ECD and UV Detector on 5 Catecholamines

Norepinephrine Epinephrine Normetanephrine Dopamine Metanephrine
ECD Detection Limits [pg] 31.1 61.7 42.6 44.8 70.4
UV Detector Detection Limits [pg] 4170 7450 9720 6680 10500
Sensitivity Comparison UV Detector / ECD 134 121 228 149 149

About the Author

Chromatography Group