Combined analysis of plastic particles by Infrared and Raman microscopy

June 15, 2021

Introduction

 

Infrared (IR) microscopy and Raman microscopy are excellent methods for the analysis of molecular structure from molecular vibrations, each of these analytical method has its own distinct merits. On the one hand, there is a large database of IR spectra for organic substances, a major attraction to users when identifying unknown foreign materials. IR spectroscopy is also useful for the analysis of proteins that are more challenging by Raman spectroscopy. On the other hand, Raman spectroscopy is well suited to the analysis of organics, inorganics and carbon-based materials, which can be more arduous by IR spectroscopy.

By combining both techniques it is possible to obtain much more detailed information than from either analysis by itself. The accuracy of this comparative analysis has been dramatically improved by the newly developed IQ Frame, an accessory in which IR and Raman measurement can be performed at exactly the same sample position (Fig. 1). To demonstrate this, in this application note, the component distribution of a sample sprayed with plastic particles was sequentially analyzed using both IR and Raman microscopy.

Fig.1 IQ Frame

Experimental

Measurement conditions
[IR microscopic system]
Mode:  Transmission mode (16x Cassegrain objective)
Resolution:   8 cm-1
Accumulation:  x1
Measurement area:   400 µm x 400 µm (12.5 μm x 12.5 µm/pixel)
Measurement points:   32 x 32 points (1024 points, measurement time: approx. 10 seconds)

 

Fig.2 IR microscope system (IRT-7200 + FT/IR-6700)

[Raman microscopic system]
Objective lens:   20x
Exposure time:   0.5 secs
Accumulation:   x1
Measurement area:   348 µm x 368 µm (4 µm x 4 µm interval)
Measurement points:   88 points x 93 points (8184 points, measurement time: approx. 70 min.)

 

Fig.3 Raman microscope system (NRS-4500)

Results

[IR microscopy]
As a result of analyzing the component distribution of plastic particles using an FT/IR7200 high-speed imaging linear-array IR microscope (Fig. 2), protein was found in addition to four plastics: polyethylene terephthalate (PET), polyethylene (PE), polypropylene (PP), and polystyrene (PS) (Fig. 4).

Fig.4 Result of IR component analysis

[Raman microscopy]
Imaging measurement was performed at the same position as by IR microscopy in Fig. 4 with a high-speed imaging Raman microscope NRS-4500QRI using EMCCD (Fig. 3). IQ Frame was used to transfer the measurement positions between the Raman and IR microscope stages. The component distribution of the plastic particles was re-analyzed and carbon was found in addition to the four plastics previously identified by IR: PET, PE, PP and PS (Fig. 5). In Fig. 4, protein was detected in addition to the four plastics, whereas in Fig. 5, carbon was detected without protein detection. These results are obtained because Raman spectroscopy has low detection sensitivity for protein, but it is good for carbon analysis.

Fig.5 The result of Raman component analysis

[Overlay of the imaging results of IR and Raman analysis]
As a result of the sample imaging shown in Fig. 4 and Fig. 5, only five components could be detected in the plastic particle sample by either IR or Raman spectroscopy separately. However, the results of combined analysis using IQ Frame indicated that six different components could be detected and imaged. In order to visualize the complete analytical information, both sets of imaging data were overlaid (Fig. 6), this showed the protein detected by IR analysis together with the five other components, PET, PE, PP, PS and carbon detected by Raman analysis. These results demonstrate that using the strong points of both IR and Raman analysis can be combined to be more useful than a single analytical technique alone.

Fig.6 Overlaid image of IR and Raman

About the Author

Spectroscopy Group