Analysis of Phenols in Tap Water by Triple Quadrupole LC/MS/MS

April 20, 2023

Introduction

Phenols are used as preservatives, and the phenols that flow into raw tap water from factory wastewater react with chlorine used as a disinfectant and produce chlorophenols. Chlorophenols give drinking water bad smell and taste, so they are one of the water quality standards defined by the Water Supply Act in Japan from the viewpoint of preventing the generation of bad smell. Solid-phase extraction-liquid chromatography-mass spectrometry has been adopted as the measurement method.

The standard value for phenols is 0.005 mg/L or less for the total of phenol, 2-chlorophenol, 4-chlorophenol, 2,4-dichlorophenol, 2,6-dichlorophenol, and 2,4,6-trichlorophenol (converted to phenol). In addition, the quantification limit is 0.0005 mg/L, one-tenth of the standard value, and the reproducibility is 20 % or less.

In this report, we examined the measurement of phenols in tap water by LC/MS/MS using a SCIEX triple quadrupole mass spectrometer (Tripe Quad™ 3500) as a detector, and also conducted recovery tests by solid-phase extraction (SPE).

LC-4000 HPLC system (left) and Tripe Quad™ 3500 triple quadrupole mass spectrometer (right)

Experimental

[HPLC Conditions]
Column:   InertSustain C18 HP (2.1 mm I.D. x 100 mm L, 3 µm)
Eluent A:  Water
Eluent B:  Methanol
Gradient:  A/B = 60/40 (0.00 min.)
–> 10/90 (6.00 min.)
–> 10/90 (8.00 min.)
–> 60/40 (8.05 min.) , 1 cycle; 12 min.
Flow  rate:  0.4 mL/min.
Column temp.:   40 ºC
Injection volume:   50 µL
Standard:  Mixture of phenol, 2-chlorophenol, 4-  chlorophenol, 2,4-dichlorophenol, 2,6-  dichlorophenol, and 2,4,6-trichlorophenol in 20 % methanol (4 µg/L, 10 µg/L, 20 µg/L, 40 µg/L, 80 µg/L each)

[MS Conditions]
Ionization:   APCI, Negative
Temperature:   400 ºC
MRM:
Q1/Q3 (93.0/65.0) for phenol
Q1/Q3 (126.9/90.9) for chlorophenol
Q1/Q3 (160.8/124.9) for dichlorophenol
Q1/Q3 (194.8/158.7) for trichlorophenol

Keywords

phenols, water quality standard, SPE, triple quadrupole mass spectrometer, LC/MS/MS

Results

Figure 1 shows the measurement result of standard mixture of 6 phenols (4 µg/L each). Since the test water (tap water) is finally concentrated by 50 times with SPE as described below (refer to Figure 3), the range of the calibration curve was also set to the concentration after 50 times concentration.

The concentrations per one component calculated from the reference values are shown in Table 1. In this experiment, the minimum and maximum concentrations of each component were set to 4 µg/L and 80 µg/L, respectively, and calibration curves were prepared at a total of five points (Figure 2). All components showed good linearity with more than 0.9999 of correlation coefficients (r) in the range of 4 µg/L to 80 µg/L.

Fig. 1   Chromatograms of standard mixture of 6 phenols (4 µg/L each)
1: Phenol, 2: 2-Chlorophenol, 3: 4-Chlorophenol, 4: 2,6-Dichlorophenol, 5: 2,4-Dichlorophenol, 6: 2,4,6-Trichlorophenol

Table 1   Concentration range of calibration curves calculated from standard values of 6 phenols

Concentration of test water Concentration after concentrating by 50 times
Standard values Sum of 6 components 0.005 mg/L 0.25 mg/L
1 component 0.833 µg/L 41.6 µg/L
One-tenth of standard values Sum of 6 components 0.0005 mg/L 0.025 mg/L
1 component 0.0833 µg/L 4.16 µg/L
Range of calibration curves Sum of 6 components 0.0005 – 0.01 mg/L 0.025 – 0.5 mg/L
1 component 0.0833 – 1.66 µg/L 4.16 – 83.3 µg/L

Fig. 2   Calibration curves of 6 phenols

Tables 2-1 and 2-2 show the reproducibilities of 6 phenols at 4 µg/L (one-tenth of the standard value) and at 40 µg/L (standard value), respectively. At both concentrations of all of the components, good reproducibilities of retention time and peak area were obtained with less than 0.1 % and less than 3.5 %, respectively.

Table 2-1   Reproducibilities of  retention time and peak area for 6 phenols at 4 µg/L (n = 5)

Number of Injection Phenol 2-Chlorophenol 4-Chlorophenol 2,6-Dichlorophenol 2,4-Dichlorophenol 2,4,6-Trichlorophenol
tR [min.] Peak area tR [min.] Peak area tR [min.] Peak area tR [min.] Peak area tR [min.] Peak area tR [min.] Peak area
1 2.036 38279 3.327 9384 3.784 16775 4.432 10165 5.093 109970 5.987 7074
2 2.034 37435 3.328 9071 3.783 17482 4.435 10316 5.095 110363 5.988 7088
3 2.035 36444 3.327 9140 3.783 17132 4.434 9931 5.095 109513 5.991 7180
4 2.036 34962 3.330 9292 3.784 17096 4.436 10274 5.097 107687 5.993 7091
5 2.039 36556 3.333 9161 3.787 17683 4.439 10006 5.099 109612 5.992 6998
Average 2.036 36735 3.329 9210 3.784 17234 4.435 10138 5.096 109429 5.990 7086
SD 0.0018 1238 0.0024 126 0.0018 355 0.0026 167 0.0023 1029 0.0025 65
RSD [%] 0.09 3.37 0.07 1.37 0.05 2.06 0.06 1.65 0.04 0.94 0.04 0.91

Table 2-2   Reproducibilities of  retention time and peak area for 6 phenols at 40 µg/L (n = 5)

Number of Injection Phenol 2-Chlorophenol 4-Chlorophenol 2,6-Dichlorophenol 2,4-Dichlorophenol 2,4,6-Trichlorophenol
tR [min.] Peak area tR [min.] Peak area tR [min.] Peak area tR [min.] Peak area tR [min.] Peak area tR [min.] Peak area
1 2.036 355478 3.328 96217 3.784 180911 4.434 109138 5.096 1155512 5.992 73093
2 2.037 346721 3.329 96440 3.784 177898 4.436 108470 5.097 1138848 5.993 74427
3 2.039 353827 3.334 96966 3.788 177861 4.440 106588 5.100 1162063 5.995 73903
4 2.039 348322 3.336 95948 3.790 178218 4.440 108454 5.101 1138471 5.995 75093
5 2.038 359943 3.333 96558 3.787 180877 4.440 105689 5.100 1135895 5.994 74405
Average 2.038 352858 3.332 96426 3.786 179153 4.438 107668 5.099 1146158 5.994 74184
SD 0.0014 5391 0.0033 381 0.0026 1595 0.0027 1458 0.0020 11814 0.0013 742
RSD [%] 0.07 1.53 0.10 0.40 0.07 0.89 0.06 1.35 0.04 1.03 0.02 1.00

Figure 3 shows the pretreatment procedure for the concentration of test water (tap water) by SPE. SPE cartridge used was InertSep Slim-J PLS-3 filled with nitrogen-containing methacrylate-styrene divinylbenzene polymer  (manufactured by GL Sciences).

Fig. 3   Procedure for concentration of test water by solid phase extraction (50 times concentration)
* If residual chlorine is included, add sodium ascorbate (0.01-0.02 g/mg Cl2 as a guide). 5 mg of sodium ascorbate was added in this experiment.

Sample pretreatment was carried out according to the procedure shown in Figure 3 for three samples: tap water only, tap water containing standard mixture of 6 phenols (4 µg/L each, one-tenth of the standard value; 0.08 µg/L in test water), and tap water containing standard mixture of 6 phenols (40 µg/L each, standard value; 0.8 µg/L in test water).

Figure 4 shows the results of each sample. In the tap water, phenol and 2,6-dichlorophenol were detected at concentrations lower than one-tenth of the standard value. As shown in Table 3, good recovery rates of 85 % to 107 % were obtained for all phenols detected in both tap water samples containing 4 µg/L and 40 µg/L of standard mixture.

Fig. 4   Chromatograms of tap water added at 4 or 40 µg/L of standard mixture in recovery test
1: Phenol, 2: 2-Chlorophenol, 3: 4-Chlorophenol, 4: 2,6-Dichlorophenol, 5: 2,4-Dichlorophenol, 6: 2,4,6-Trichlorophenol

Table 3   Recovery rates of 6 phenols in tap water by SPE

Recovery rate [%] Phenol 2-Chlorophenol 4-Chlorophenol 2,6-Dichlorophenol 2,4-Dichlorophenol 2,4,6-Trichlorophenol
Tap water with 4 µg/L 100.4 106.6 100.1 97.0 106.3 94.6
Tap water with 40 µg/L 85.2 94.5 94.9 89.4 90.6 86.0

About the Author

Chromatography Group