Rapid FTIR Quantitative Analysis of Trans-Fatty Acid at Extremely Low Concentrations using a Liquid Cell

September 30, 2020

Introduction

Trans-fatty acids in food are becoming a major concern due to their detrimental affects on human health. A method for the rapid quantitative analysis using FTIR with temperature controlled ATR has been suggested1) and an official method using FTIR has been proposed by both the AOAC (American Organization of Analytical Chemists) and AOCS (American Oil Chemists’ Society) due to its extremely short analysis time, within 1–2 minutes, which is much shorter than that required by a GC method (Gas Chromatography, which also requires extensive sample preparation). The official method described by the AOCS, suggest that the limit of quantitation (LOQ) using the ATR method is around 1.0%1), this was also demonstrated in our FTIR Application data under the same experimental conditions specified by the AOCS.

According to “The Guideline for Disclosure of Information on Trans-fatty acid Content” issued by the Consumer Affairs Agency of Japan, “Zero” may be shown on the labeling if the measured content of trans-fatty acid in food is less than 0.3% w/w or 0.3% w/v (for beverages). In other words, 0.3% is the required limit of quantitation (LOQ). Similar requirements are found in other countries (USA: < 0.5 g / meal, Taiwan: < 0.3 g / 100 g, Korea or South America: < 0.2 g / meal) and accordingly, a rapid and precise method is required for the quantitative analysis of trans-fatty acid content in food to limits below 1.0%. The method for quantitation of trans-fatty acid described in the Guideline uses GC (AOCS Ce1h-05 or AOAC996.06), or other methods that have the equivalent performance to this method. The analysis using GC not only requires extensive sample preparation before measurement, including removal of methylated fat using BF3, but also requires more than 1 hour for the separation. This report describes the quantitative analysis of trans-fatty acids to levels below 1.0% using an FTIR transmission method, demonstrating an equivalent result to the GC method and can be obtained in an extremely short time.

Experimental

The trans-fatty acid peak at 966 cm-1 is used for the quantitative analysis (Figure 1). This is also adopted in the official methods, such as the ATR method1).

Figure 1. IR spectrum of cis-fats and trans-fats(ATR)

A calibration curve was created with 5 standard levels of trielaidin, which were prepared by adding trielaidin (green color spectrum in Figure 1), an isomer of trans-triolein, to cis-triolein (orange color spectrum in Figure 1) which contains only cis-type.

Measurement Conditions
Instrument
Detector DLaTGS Resolution 4 cm-1
Accumulation 64 times Apodization Triangle
Temperature 25°C (ambient temperature)
Mode Transmission (method of solution)
Cell Sealed liquid cell *NaCl (thickness : 0.1 mm)
Standard Sample Triolein, Trielaidin (0.05, 0.1, 0.2, 0.5, 1.0%)
Peak Calculation Area 945 – 990 cm-1
* A sealed liquid cell with KBr windows can be also used.
Please contact local JASCO distributor if the cell needs to be thermostatted.

Figure 2. Sealed liquid cell

Keywords

trans fatty acids, trans fats, FTIR, transmission, ATR, LDL, HDL, cholesterol

Results

Transmission spectra of 5 standard levels of trielaidin (0.05, 0.1, 0.2, 0.5, 1.0%) are displayed in Figure 3.

Figure 3. Trans-fats peak at 966 cm-1

The correlation coefficient (0.9998) for the calibration curve shows a good linearity for the trans-fatty acid concentration using the peak height or area around 966 cm-1 (Figure 4).

Figure 4. Calibration curve

Since the S/N (Peak-to-Peak) for the standard level at 0.05% is about 15:1, which is much better than the required 10:1 normally accepted as the limit of quantitation (LOQ) (Figure 5), this method is considered sensitive enough for the quantitative analysis of samples with low concentrations, down to 0.05%.

Figure 5. Spectrum of trans-fats of 0.05%

Conclusion

The result shown indicates that the quantitative analysis of 0.05 – 1.0% trans-fatty acid can be achieved using FTIR transmission measurement. This method has the distinct advantages of little or no sample preparation and is much faster compared with the GC method. However, the cell has to be washed between measurements, making it a little less convenient than the ATR method. Therefore, we recommend that the ATR method is used for measurement to LOQs of 1% and the transmission method is used for sample measurement to LOQs below 1.0%.

References

1) AOAC Official Method 2000.10, AOCS Official Method Cd 14d-99