Rapid Separation of a Tuberculosis Medication using UHPLC

December 11, 2020

Introduction

High speed HPLC separations have become increasingly important in a variety of different fields including pharmacology, food science, and agrochemistry. In general, conventional HPLC separations have been performed on columns packed with a stationary phase composed of 5 µm particles. By contrast, UHPLC system uses shorter columns packed with smaller sub-2 µm particles. This article details the analysis of a tuberculosis treatment composed of three compounds (rifampin ,isoniazid, and pyrazinamide) with a method outlined by USP to compare the performance of UHPLC with conventional HPLC.

LC-4000 UHPLC system

Experimental

The conventional and rapid sample analyses described below were made using an integrated UHPLC system configured for binary gradient with UV-visible detection.

Chromatographic conditions

Mobile Phase A: 0.01 M Na2HPO4/CH3CN (96/4) (pH 6.8)
Mobile Phase B: 0.01 M Na2HPO4/CH3CN (45/55) (pH 6.8)
Wavelength 238 nm
Sample 0.16 mg/L Rifampin, 0.08 mg/L Isoniazid, 0.43 mg/L Pyrazinamide, (in 0.01M Na2HPO4/CH3OH (96/4))

Results

Figure 1 shows chromatogram of medicines using conventional HPLC. A flow rate of 1.5 mL/min was used with a CrestPak C18S column, 5 µm particle size (4.6 mm i.d. X 150 mm L). The analysis time was 9 min.

Figure 1 Chromatogram of medicines using column (4.6 mm I.D. x 150 mm) packed with 5 µm particles in diameter (1: Rifampin, 2: Isoniazid, 3: Pyrazinamide)

Figure 2 shows chromatogram of medicines using UHPLC. A flow rate of 0.8 mL/min, nearly half that of the HPLC separation, was used for elution of the sample using an X-PressPak C18S column with 2 µm particles (2.1 mm i.d. X 50 mm L). This resulted in an analysis time of 1.3 min, approximately 7 times faster than the HPLC separation. The gradient elution program of 1.5 min is 10 times shorter than that for HPLC.

Figure 2 Chromatogram of medicines using column (2.1 mm I.D. x 50 mm L.) packed with 2 µm particles in diameter (1: Rifampin, 2: Isoniazid, 3: Pyraziamide)

Table 1. Comparison between UHPLC and conventional HPLC.

Pyrazinamide (Peak #2) HPLC UHPLC UHPLC/HPLC
Analysis Time (min) 8 1.3 0.163
Solvent (mL/analysis) 30 1.6 0.053
Peak Height (mAU) 337 378 1.122
Injection Volume (µL) 10 1 0.1

Conclusion

We have demonstrated that UHPLC system offers rapid separations, up to 10 times faster than that of conventional HPLC, with improved sensitivity and efficiency whilst reducing mobile phase solvent use. This more efficient technique is also completely scaleable from traditional HPLC methods providing an attractive option for a wide range of samples.

About the Author

Chromatography Group