Optical Rotation Measurement for High-concentration Acidic Sample Solutions

January 7, 2026

Introduction

Key Points
The P-4000 polarimeters offer high durability for acidic sample measurements because of an acid-resistant coating on the sample compartment wall.

In certain optical rotation tests described in the United States pharmacopeia (USP) and the Japanese pharmacopoeia (JP), sample preparation requires the use of acidic solutions such as 5 N hydrochloric acid (USP) or 6 mol/L hydrochloric acid (JP). Because optical rotation is temperature-dependent, it is important to monitor the temperature during measurements. The P-4000 series polarimeters are equipped with a temperature sensor that can be inserted directly into the cell. This cell sensor has an acid-resistant coating, allowing direct measurement of the temperature of acidic sample solutions. To protect the instrument when measuring acidic liquid samples, the P-4000 series incorporates the following features:

1.Acid-resistant coating on the sample compartment wall
2.Quartz window preventing acidic gas from entering the instrument*
3.Ventilation of acidic gas*
4.Connection to a ventilation duct**
* Standard feature only for the P-4200 and P-4300
** Optional for the P-4200 and P-4300

This application note describes the measurement of L-alanine in 6 mol/L hydrochloric acid.

Fig. 1 P-4200 Polarimeter

Experimental

Sample
L-alanine (research reagent)
In accordance with the monograph described in JP XVIII1) (after drying, 2.5 g, 6 mol/L hydrochloric acid TS, 25 mL, 100 mm), the sample was prepared by dissolving dried L-alanine in a 6 mol/L hydrochloric acid solution.

System
Instrument: P-4200 polarimeter
Cell: RQC-450 8.5-mm rectangular quartz cell (path length: 100 mm)

Measurement conditions
Light source: Na lamp
Wavelength: D-line
Integration time: 5 sec
Number of repeats: 1

Keywords

Polarimeter, acid resistant, 6 mol/L hydrochloric acid, L-alanine

Results

Table 1 shows the measurement results for the specific rotation of L-alanine. The acid-resistant cell sensor enables accurate monitoring of the sample temperature, even for a strong acid solution of 6 mol/L hydrochloric acid, allowing precise measurement of the optical rotation.

Table 1   Results for L-alanine

Specific rotation Temperature during measurement / ºC
+14.824 20.4

Conclusion

The P-4000 series polarimeters allow accurate measurement of optical rotation for highly acidic sample solutions, such as 6 mol/L hydrochloric acid, while directly monitoring the sample temperature. Additionally, these polarimeters exhibit high resistance to acidic samples, enabling reliable operation while maintaining the instrument in optimal condition.

References

1.Ministry of Health, Labour and Welfare: June 7, 2021, MHLW Ministerial Notification No. 220, “The Japanese Pharmacopoeia 18th edition”, (2021).