Introduction
Key Points
The use of short-path-length cells allows measurements of samples with a large optical rotation.
JASCO offers cells with path lengths of 10 mm, 20 mm, 50 mm and 100 mm.
Some types of pure liquid* cannot be measured with a polarimeter and a normal 100-mm-path-length cell because their optical rotation exceeds 90 degrees. JASCO offers cells with short path lengths that enable measurements of samples with a large optical rotation. This application note describes optical rotation measurements for orange oil with a 50-mm-path-length cell.
*Pure liquid: a liquid substance composed of only one type of molecule

Fig. 1. P-4200 polarimeter
Experimental
Sample
Orange oil (research use)
System
Instrument: P-4200 polarimeter (Fig. 1)
Cell: RQC-451 8.5-mm rectangular quartz cell (path length: 50 mm) (Fig. 2)

Fig. 2. Rectangular cell
Parameters
Light source: Na lamp
Wavelength: D line
D.I.T.: 5 sec
Repetition: 1 time
Keywords
Polarimeter, short path length, pure liquid, orange oil
Results
The measurement results are shown in Table 1. The blank measurement was performed using a dry empty cell.1) The use of a 50-mm-path-length cell allowed the optical rotation for orange oil to be measured without dilution, as the measured value remained within 90 degrees.
Table 1 Results
| Optical rotation / ° | Temperature / ºC |
| 49.107 | 20.4 |

Fig. 3. Measurement screen
Top: The current optical rotation and temperature are displayed in the monitor.
Bottom: Sample names can be added to the list for future measurement.
Conclusion
Using a short-path-length cell, samples with a large optical rotation can be measured without dilution.
References
1.United States Pharmacopeial Convention: “USP-NF 2024 Issue 3”, (2024).
2.Ministry of Health, Labour and Welfare: June 7, 2021, MHLW Ministerial Notification No. 220, “The Japanese Pharmacopoeia 18th edition”, (2021).