JASCO unique rectangular cell with superior usability
Novel sampling cell was developed based on the customer feedback and needs.
(US patent: US D925,381, Europe patent: DM/205719, UK patent: GB 82057190001000, China patent: ZL201930223409.3)
Efficient pouring and flushing
RQC cells are designed for easy, bubble-free sample pouring and thorough flushing.
Superior temperature control
RQC cells quickly reach the set temperature due to the large contact area between cell jacket and cell.
Small-volume sampling
Minimum sample volume of 3 mm RQC cell with 100 mm optical pathlength is approx. 1 mL.

RQC-453 Rectangular quartz cell (3 mm × 3 mm × 100 mm)

RQC cell jacket
Plenty cell lineup
Each sampling cell is used with the dedicated cell jacket.
| Model | P-4100 | P-4200 | P-4300 |
| Rectangular quartz cell | |||
| RQC-450 | 8.5 mm × 8.5 mm × 100 mm | ||
| RQC-451 | 8.5 mm × 8.5 mm × 50 mm | ||
| RQC-452 | 8.5 mm × 8.5 mm × 10 mm | ||
| RQC-453 | -* | 3 mm × 3 mm × 100 mm | |
| RQC-454 | -* | 3 mm × 3 mm × 50 mm | |
| RQC-455 | -* | 3 mm × 3 mm × 20 mm | |
| Cylindrical quartz cell | |||
| J/39.105/Q/100/12 | φ10 mm × 100 mm | ||
| J/39.105/Q/50/12 | φ10 mm × 50 mm | ||
| J/39.35/Q/100/6 | -* | φ3.5 mm × 100 mm | |
| J/39.35/Q/50/6 | -* | φ3.5 mm × 50 mm | |
| J/39.35/Q/10/6 | -* | φ3.5 mm × 10 mm | |
| Cylindrical glass cell | |||
| J/39.105/PX/100 | φ10 mm × 100 mm | ||
| J/39.105/PX/50 | φ10.5 mm × 50 mm | ||
| J/39.105/PX/10 | φ10.5 mm × 10 mm | ||
| SUS demountable cell | |||
| PAC1-100 | φ10 mm × 100 mm | ||
| PAC3-100 | -* | φ3.5 mm × 100 mm | |
*3 × 3 mm rectangular cell and φ3.5 mm cylindrical cell do not work with P-4100.

RQC-450 Rectangular quartz cell (8.5 mm × 8.5 mm × 100 mm)

J/39.35/Q/50/6 Cylindrical quartz cell (φ3.5 mm × 50 mm)

J/39.105/PX/10 Cylindrical glass cell (φ10.5 mm × 10 mm)

PAC1-100 SUS demountable cell (φ10 mm × 100 mm)
Dedicated accessories for P-4000 Series
SHP-403 Peltier sipper
SHP-403 allows sample loading from the front nozzle, followed by measurement and flushing. It saves time and effort in mounting and dismounting the cell.

TSR-411 Thermostatted plate
TSR-411 can preheat the sampling cell before the measurement, reducing the overall measurement time, including the heating after cell placement.

QPA-412/413/414/415
Practical standard quartz control plate
QPA is used for calibrating the instrument. JASCO also offers the quartz control plate with calibration certificate compliant with ISO/IEC 17025 compliance.

Optical filter
Optical filters must be changed according to the measurement wavelength. P-4300 has the automatic filter changer as standard.