Analysis of bicalutamide listed in the Japanese Pharmacopoeia 18th Edition

November 27, 2025

Introduction

Bicalutamide is a newly listed drug in the 18th edition of the Japanese Pharmacopoeia (JP) for the treatment of prostate cancer. This drug inhibits the binding of male hormones (androgens) to androgen receptors, thereby suppressing the growth of cancer cells and exerting an anti-tumor effect.

In this report, we present the results of purity and assay on bicalutamide in accordance with the Japanese Pharmacopoeia, 18th Edition.

LC-4500 series HPLC system

Experimental

<Instruments>
Pump: PU-4580
Degassing unit: DG-4580
Low pressure gradient unit: LG-4580
Autosampler:  AS-4550
Column oven:  CO-4060
UV detector:  UV-4570

<Conditions>
Column:   InertSustain C18 (4.0 mmI.D. x 250 mmL, 5 µm)
Eluent A:  0.1 % Phosphoric acid/acetonitrile (19/1)
Eluent B:  Acetonitrile/0.1 % phosphoric acid (19/1)
Gradient:  A/B = 92/8 (0.0 min) -> 67/33 (20.0 min) -> 50/50 (40.0 min) -> 50/50 (47.0 min) -> 92/8 (47.1 min), 1cycle; 57 min
Flow rate:  1.0 mL/min
Column temp.:   50 ºC
Preheat coil:  0.25 mmI.D. x 1000 mmL
Wavelength:   210 nm
Inj. volume:   10 µL
Standard:  Bicalutamide (The dissolving and diluting solvent   is a mixture of water/acetonitrile/phosphoric acid   (1000/1000/1).)

<Structure>

Bicalutamide

Keywords

Japanese Pharmacopoeia, bicalutamide, UV detector

Results

An overview of the system suitability for bicalutamide purity and assay is shown in Table 1.

Table 1   Overview of system suitability for bicalutamide purity and assay

Test Contents
Purity Detectability When the standard solution for purity (1 mg/L) is diluted to 0.5 mg/L and the test is performed under the above operating conditions on this solution, the SN ratio for the bicalutamide peak is ≥10.
System performance When the test is performed on 10 µL of the standard solution for purity under the above operating conditions, the number of theoretical plates and the bicalutamide peak symmetry factor are ≥10000 and ≤1.5, respectively.
System reproducibility When the test is repeated 6 times with 10 µL of the standard solution for purity under the above operating conditions, the relative standard deviation of the bicalutamide peak area is ≤5.0 %.
Assay System performance When the test is performed with 10 µL of the standard assay solution (200 mg/L) under the above operating conditions, the number of theoretical plates and the bicalutamide-peak symmetry factor are ≥10000 and ≤1.5, respectively.
System reproducibility When the test is repeated 6 times with 10 µL of the standard assay solution under the above operating conditions, the relative standard deviation of the bicalutamide peak area is ≤1.0%.

(1) Purity

Figure 1 shows the results of the detectability. For the standard solution for purity diluted to 0.5 mg/L, the SN ratio for the bicalutamide peak was 110.7, which meets the criterion of ≥10.

Fig. 1   Chromatogram of bicalutamide for detectability (0.5 mg/L)

Figure 2 shows chromatograms of bicalutamide standard solutions (n = 6), and Table 2 shows the results for peak area reproducibility based on the measured system performance and system reproducibility for purity. The number of theoretical plates was 208,383 (criterion: ≥10000), the symmetry factor was 0.98 (criterion: ≤1.5), and the relative standard deviation of the peak area was 0.12% (criterion: ≤5.0%), all of which met the specified criteria.

Fig. 2   Chromatogram of bicalutamide standard solutions (1 mg/L, n = 6)

Table 2   Bicalutamide peak area    reproducibility (n = 6)

Injection number Peak area
1 27505
2 27505
3 27521
4 27433
5 27491
6 27521
Ave. 27496
SD 33
RSD[%] 0.12

(2) Assay

Figure 3 shows chromatograms of bicalutamide standard solutions (n = 6), and Table 3 shows the peak area reproducibility based on the measured system performance and system reproducibility for the assay. The number of theoretical plates was 206,269 (criterion: ≥10000), the symmetry factor was 0.97 (criterion: ≤1.5), and the relative standard deviation of the peak area was 0.28% (criterion: ≤1.0 %), all of which all of which met the specified criteria.

Fig. 3   Chromatograms of bicalutamide standard solutions (200 mg/L, n = 6)

Table 3   Bicalutamide peak area       reproducibility (n = 6)

Injection number Peak area
1 5735272
2 5722046
3 5719084
4 5698454
5 5699959
6 5701384
Ave. 5711484
SD 15835
RSD[%] 0.28

Conclusion

We evaluated the system suitability for purity and assay on bicalutamide, which is newly listed in the 18th edition of the Japanese Pharmacopoeia. As shown in Table 4, all of the evaluation results met the specified criteria for the Japanese Pharmacopoeia.

Table4   Results of system suitability evaluation for bicalutamide purity and assay

Test Test item Criteria Result Judgement
Purity Detectability SN ratio ≥10 110.7 Passed
System performance Number of theoretical plates ≥10,000 208,383 Passed
Symmetry factor ≤1.5 0.98 Passed
System reproducibility Relative standard deviation of peak area ≤5.0% 0.12 % Passed
Assy System performance Number of theoretical plates ≥10,000 206,269 Passed
Symmetry factor ≤1.5 0.97 Passed
System reproducibility Relative standard deviation of peak area ≤1.0% 0.28 % Passed

About the Author

Chromatography Group